product description
Specifications
PARAMETER |
SPECIFICATION |
Internal size(W*D*H) |
1220*600*630mm |
External size(W*D*H) |
1380*760*2100mm |
Tested Opening |
Safety Height 200mm (8’’) |
Max Opening |
440mm (17″) |
Internal Work Area, space |
0.73m3 |
Inflow Velocity |
0.53±0.025m/s |
Down Flow Velocity |
0.33±0.025m/s |
Exhaust Volume |
465m³/h (237cfm) |
Air Supply Filter |
ULPA filter, efficiency>99.9995% for particle size at 0.12μm |
Exhaust Filter |
ULPA filter: air filter: 0.12μm efficiency 99.9995%; ULPA filter: first exhaust filter: 0.12μm efficiency 99.9995%; ULPA filter: secondary exhaust filter: 0.3μm efficiency 99.9995% |
Alarm |
Abnormal airflow velocity; filter replacement, Front window at unsafe height; High filter pressure alarm |
Noise |
EN12469≤58dB/NSF49≤ 61 dB |
Water &Gas Tap |
Water Tap*1, Gas tap*1 |
UV lamp |
30W*1. Emission of 253.7nanometers for most efficient decontamination UV life indicator, UV timer |
LED Lamp |
16W*2 |
Illumination |
≥1000 lux |
Consumption |
900W |
Waterproof Socket |
Two; total load of two sockets: ≤500W |
Display |
LCD Display |
Control System |
Microprocessor |
Humidity and Temperature |
Humidity and temperature can be displayed on the control panel |
Material |
Work Area: 304 stainless steel; Main Body: Cold-rolled steel with anti-bacteria powder coating |
Front Window |
Motorized, two-layer laminated toughened glass≥6mm, Anti-UV. |
Caster |
Foot master Caster |
Power Supply |
AC220V±10%,50/60HZ; 110V±10%, 60HZ |
Standard Accessory |
Base stand, Drain valve, waterproof sockets*2, Remote control, UV lamp*2 and LED Lamp, Water tap, Gas tap |
Optional Accessory |
Armrest, Formalin fumigation sterilizer, infrared sterilizer, Airflow tester |
Gross Weight |
371kg |
Package Size(W*D*H) |
1540*910*2280mm |