V-AC9801

Vet Electrolyte Analyzer

product description

Technical Specifications

Model

AC9801

Measuring item

K+、Na+、Cl-、Ca、pH

Calculating item

nCa、TCa

Measuring parameter

Measuring

item

Measuring scope

(mmol/L)

Resolving power

(mmol/L)

Repeatability

error

(K+)

0.50~15.00

0.01

CV≤1.0%

(Na+)

30.0~200.0

0.1

CV≤1.0%

(Cl-)

30.0~200.0

0.1

CV≤1.0%

(Ca2+)

0.10~5.00

0.01

CV≤1.0%

pH

4.0~9.0

0.01

CV≤1.0%

Measuring time

Measuring time: 25s

Time for sampling, measuring, washing and printing: 40s

Sample size

100μL(manual positioning 60μL)

Data storage

10000,It will refresh automatically if it is full

Communication interface

232 interface

Display

240×64 LCD

Printer

58mm thermal printer

Power

supply

AC220V±22V 50Hz 25W

Size

380mm×290mm×310mm

Weight

6Kg

Measuring method

ISE

Measuring

condition

Temperature: 10°C ~32°C  

Relative humidity ≤85%